Place the cell on the holder with the screen side facing down.
NOTE: There is a small location hole drilled in the holder. This hole mates with a guide pin on the bottom rear of the cell block. The hold in the cell block holder must align with the guide pin on the cell block.
STEP 1: Remove power from instrument.
STEP 2:Remove the old sensor (if installed) from the analyzer.
STEP 3: Purge the analyzer at approximately 1 SCFH flow rate with N2 (or applicable sample gas with the sensor holder removed.)
STEP 4: Remove sensor from double bag storage.
STEP 5: Remove sensor shorting button.
STEP 6: Place the sensor inside the cell block.
STEP 7: Hold the sensor up into the block while positioning the cap under the sensor. Vigorously push the cell holder with the sensor into the cell block, thus puncturing the protective film / seal cover.
STEP 8: Push the gate on the cell block down so that the slots on the side of the gate engaged the locating screws on the side of the block. This forces the holder into position and forms a gas-tight seal.
STEP 9: Purge system with sample or zero gas.
STEP 10: Power up.
Your Insta Trace sensor will typically reach a stable low value within 30 minutes or less of installation.